Skip to main content

Table 6 Comparison of averaged insertion torques from experiments and predictions

From: Oblique Cutting Based Mechanical Model for Insertion Torque of Dental Implant

Thread number

Group 1

Group 2

Group 3

Measured

(N·cm)

Pred

(N·cm)

Measured

(N·cm)

Pred

(N·cm)

Measured

(N·cm)

Pred

(N·cm)

1

1.899

1.669

2.256

0.693

1.036

2.244

1.167

3.158

2.693

2

4.872

4.559

5.270

2.237

4.693

6.012

5.394

7.061

7.214

3

9.805

10.272

8.284

4.693

7.125

9.779

8.171

10.475

11.735

4

12.050

12.197

11.298

8.693

12.362

13.547

12.617

15.392

16.256

5

15.874

13.893

14.312

13.369

14.639

17.314

23.872

22.389

20.777

6

17.274

15.325

17.326

15.693

18.693

21.082

25.349

28.869

25.298

7

18.326

20.186

20.340

18.639

22.237

24.849

33.995

33.612

29.819

8

22.264

20.362

23.354

24.363

23.140

28.617

34.693

38.118

34.34

9

23.592

21.369

24.279

26.964

27.063

30.256

35.460

41.636

36.471

10

24.362

22.012

25.206

27.365

28.634

31.895

36.256

43.124

38.602

11

24.982

23.937

26.132

29.369

29.363

33.534

37.596

44.691

40.733

12

25.193

24.102

27.058

30.069

31.693

35.174

38.172

46.786

42.864

error

7.4%

12.3%

0

16.9%

10.9%

0

12.4%

8.4%

0